Xiamen Bioendo Technology Co., Ltd Official Website
  • Home
  • Products
  • Phoenix Series Intelligent Fully Automatic Endotoxin Detection Robot System
Phoenix Series Intelligent Fully Automatic Endotoxin Detection Robot System
PHOENIX Q5
Phoenix Series Intelligent Fully Automatic Endotoxin Detection Robot System
PHOENIX Q5
Bacterial endotoxin detection is a critical step in ensuring the safety of pharmaceutical and medical products. Traditional methods, though considered the gold standard, are labor-intensive, time-consuming, heavily dependent on operator expertise, and susceptible to environmental and human errors. To address these challenges, BIOENDO has developed the Phoenix Q Series Intelligent Automatic Endotoxin Detection Robot System, which integrates automated robotics with endotoxin testing. The system enables fully automated experimental workflows without human intervention, minimizing manual errors, enhancing reproducibility, and reducing overall detection costs. It is compatible with a wide range of detection methods, including kinetic turbidimetric assay, kinetic chromogenic assay, micro-kinetic chromogenic assay, recombinant cascade chromogenic assay, and the monocyte activation test (MAT). Featuring a user-friendly software interface, one-click experiment setup, and automated data recording and reporting, the Phoenix Q Series significantly improves detection efficiency and frees operators from complex manual procedures.
Catalog No. / ModelSystem
Phoenix Q5Phoenix Q5 Intelligent Fully-automatic Endotoxin Detection System
Catalog No. / ModelSystem
Phoenix Q9Phoenix Q9 Intelligent Fully-automatic Endotoxin Detection System
Note:  For other sizes/specifications not listed above, please contact our customer service team.
Email:  Eve@bioendo.com
Tel:  +86 18065928551
Technical Specifications
Phoenix Q Series
Intelligent automatic endotoxin detection robot system — Instrument parameters and ordering information.
PARAMETERSPECIFICATION
Instrument ModelPhoenix Q5  / Q9
Incubation Plate Position4
Sample Addition Channel1
Temperature Accuracy≤ ±0.5°C
Temperature Fluctuation≤ 1.0°C
Washing Microplate Residual Amount≤ 2 μL
Measurement Channel8
Stability of Display≤ ±0.005A
Absorbance Accuracy (0.000–1.000A)±0.02
Absorbance Accuracy (1.000–2.000A)±0.03
Absorbance Repeatability≤ 1%
Instrument Size (L × W × H)700 × 580 × 620 mm
Sample Volume (μL)Deviation (%)Precision CV%
50≤ ±3.0≤ 1.5
100≤ ±2.0≤ 1.0
200≤ ±1.5≤ 1.0
Ordering Information (Phoenix Q Series)
Catalog No.Product Description
Phoenix Q5Phoenix Q5 Intelligent Automatic Endotoxin Detection Robot System (without microplate washing machine)
Phoenix Q9Phoenix Q9 Intelligent Automatic Endotoxin Detection Robot System (specially designed for monocyte activation MAT method)
Automated Workflow
① Turn on ② Select experiment template  ③ Put reagent and consumables                ④ Automatic dilution and sample addition  ⑤ Incubation and oscillation   ⑥ Microplate washing (optional)                ⑦ Grab microplate away from light and reading ⑧ Automatic report generation